Model | DxF24040A DxF24040D DxF24040V | DxF32040A DxF32040D DxF32040V | DxF40040A DxF40040D DxF40040V |
---|---|---|---|
Capacity | 13 cu. ft. (368.1L) | 17.3 cu. ft. (489.9L) | 23.0 cu. ft. (653L) |
2 ml vial Capacity | 240 | 320 | 400 |
Footprint | 7.61 sq. ft. (0.71 sq. meters) | 9.0 sq. ft. (0.84 sq. meters) | 11.02 sq. ft. (1.02 sq. meters) |
Interior Dimensions | 23.1"W x 19.4"D x 51.5"H (587 x 493 x 1308 mm) | 23.1"W x 25.4"D x 51.5"H (587 x 645 x 1308 mm) | 30.6"W x 25.4"D x 51.5"H (777 x 645 x 1308 mm) |
Exterior Dimensions | 36.3"W x 31.6"D x 77.9"H (922 x 803 x 1979 mm) | 36.3"W x 37.5"D x 77.9"H (922 x 953 x 1979 mm) | 43.8"W x 37.5"D x 77.9"H (1113 x 953 x 1979 mm) |
Shipping Dimensions | 40.4"W x 38.5"D x 82.5"H (1026 x 978 x 2096 mm) | 41.8"W x 45.5"D x 82.5"H (1062 x 1156 x 2096 mm) | 49.3"W x 45.5"D x 82.5"H (1252 x 1156 x 2096 mm) |
Unit Weight | 720 lbs. (326 kg) | 713 lbs. (322 kg) | 850 lbs. (385 kg) |
Shipping Weight | -- | 830 lbs. (378 kg) | 940 lbs. (426 kg)x |
Energy Usage | 17 kW/day | 18 kW/day | 19 kW/day |
Compliance and Listings | A & D Models (UL, cUL, FDA) - V Models (CE, FDA) | ||
Included SS Shelves | 4 - Adjustable In 25 mm (1 in) Increment | ||
Shelf Capacity | 100 lbs. (45.4 kg) | ||
Refrigeration | |||
Refrigeration System | Industrial-Rated one-stage System | ||
Compressor Number | Hermetic Compressor for Low Temperature Application / 1 | ||
Compressor Capacity* | A Models - 976 W, D Models - 1026 W, V Models - 1200 W | ||
Condenser Type Number | Enhanced Finned-Tube and Forced-Air Cooled / 1 | ||
Expansion Device | Capillary Tube | ||
Evaporator Type | Cold Wall With Enhanced Heat Transfer Treatment | ||
Defrost Method | Manual Defrost | ||
Refrigerant Charge Flammability | R404A / Non-Flammable | ||
Controllers | |||
Controller Level | Top | ||
Power Switch | On-Off with Circuit Breaker | ||
Controller Type | Microprocessor Control | ||
Setpoint Security | NO | ||
Compressor Safe Guard | High Pressure Cutout Switch/High Temp Cutout Switch Current protection/Logic protection | ||
Control Sensor | Single RTD (1000 ohm Platinum RTD) | ||
Defrost Method | Manual Defrost | ||
Remote Alarm Terminals | RS485, 4-20mA, Dry Contacts | ||
Warm/Cold Alarms | Fully Adjustable | ||
Auto-Voltage Safeguard | Buck/Boost System | ||
Construction | |||
Insulation | High Density Polyurethane Foam, R-30 (ft2 °F hr/Btu) | ||
Door Seal | Silicone-Based High Performance Seal Gasket | ||
All-Direction Caster | Standard with Locks | ||
Options | Electronic Recorder and CO2 or LN2 Back-Up System or Datalogger | ||
Electrical | |||
Electrical | A Models - 120V 60 Hz, 1 P D Models - 208-230V 60 Hz, 1 P V Models - 230V 50 Hz, 1 P | ||
Temperature Range | -10 to -40 °C (14 to -40 °F) | ||
Backup Systems | Optional CO2 or LN2 systems | ||
Chart Logging | Optional | ||
Instrument Rated Current | A Models - 1 amp, D Models - 6.5 amp, V Models - 6.5 amp | ||
Building Supply Rating* | A Models - 15.0A, D & V Models - 8.0A | ||
Power Plug | A Models - 5-15P, D Models - 6-16P, V Models - Country Dependent | ||
Power Cord Length | 10 ft. (3.048 meters) | ||
Application | Storage of General (Non-Flammable) Laboratory Materials; Indoor use only | ||
Application Enviroment | Non-Corrosive, Non-Flammable, Non-Explosive, Good Air Ventilation 15 - 32° C (59 - 90° F) | ||
*With dedicated grounded circuit. Protected by circuit breaker rated for inductive loads | |||
Warranty | Two-Year Manufacturer Warranty on Parts and Labor Additional 2-years on compressor only |