FDL 115 Options and Accessories

Additional Features of FDL Series Safety Drying Chamber: More Information

  • Easy-to-change fresh air filter cartridges (replaceable) ensures constant fresh air quality
  • Symmetrical air flow with electronic monitoring system and automatic deactivation of heating
  • APT.line™ preheating technology provides optimal temperature uniformity for faster recovery times
  • Class 2 independent adjustable temperature safety device (DIN 12880) with visual alarm
  • Controller: time-segment and real-time programming
  • 60-mm-thick insulation, keeping heat dissipation to a minimum
{{dav_id_a4_alt_tag_8513}}
Options and Accessories for FDL 115 Series for Specimens Containing Solvents

The FDL 115 Safety Drying Chamber is available with a range of accessories and options to satisfy a variety of applications in norm-compliant drying applications containing solvents.

  • Access port with silicone plug (for external measuring equipment; large selection of diameters and positions (left, right, or top)
  • Calibration certificate, expanded
  • Calibration certificate, temperature
  • Cart with casters and locking brakes
  • Coil coating door flap ideal for coil coating tests
  • Door gasket made of high temperature resistant silicone for operational temperatures above 200 °C
  • Door lock
  • Measurement of air exchange rate
  • pH-neutral detergent
  • Rack
  • Rubber pads
  • Shelf (perforated, stainless steel
  • Software (Version 2 to 3, GLP, Basic, Standard and GLP DEMO editions)
  • Specimen temperature display
  • Services: Access our installation, maintenance, calibration, validation, and warranty services by taking out a maintenance service agreement
ModelFDL 115
Interior volume4.1 cu. ft. (116 L)
Outer dimensions (W x D x H)32.7 x 27 x 31.7 in. (831 x 686 x 806 mm)
Internal Dimensions(W x D x H)23.64 x 17.14 x 17.14 in. (601 x 435 x 435 mm)
Electrical data230V, 50/60Hz
Net weight of the unit (empty)198 lbs. (89.8 kg)
Temperature range 10 °C above ambient temperature to [°C]300
Wall clearance back6.3 in. (160 mm)
Wall clearance sidewise3.94 in. (100 mm)
Temperature variation at 150 °C [± K]2.5
Temperature fluctuation [± K]0.8
Heating-up time to 150 °C [min]20
Recovery time after 30 seconds door open at 150 °C [min]12
Air circulation (approx.) [x/min]20
Volumetric flow rate of exhaust air according to EN 1539 at 50 °C [ca. L/min]400
Highest permitted solvent quantity (at T-180 °C, M-100g/mol, U-40g/m3, K=0,5) [g]6.65
Nominal power [kW]2.9
Phase (Nominal voltage)1~
Energy consumption at 150 °C [Wh/h]1200
Sound-pressure level [dB(A)]57
Permitted load132 lbs. (59.9 kg)
Load per rack66 lbs. (29.9 kg)
Number of shelves (std./max.)2/5