Summary

• Requires connection to PC for operation

• Includes UV Lab Software #N4108001 and Single Cell Holder #N4103000

• Dimensions: 13.4 x 12.6 x 4.5 in (340 x 320 x 115 mm)

• Weight: 14.6 lbs.(6.6 Kg)

• Electrical: 100‐240 V 50/60 Hz

• Wavelength Range: Typical 190‐1100 nm

• Source Lamp: Xenon Flash; only active during operation

• Detector: Diode Array

• Photometric System: Single Beam Array

• Measurement 200~900 nm

• Spectral Bandwidth: 2 nm (Spectral Resolution)

• Wavelength Accuracy: ± 1 nm

• Wavelength Reproducibility: < 0.02 nm

• Photometric Range: +/‐ 3A

• Photometric Accuracy: +/‐ 0.01 AU (60 ppm

• Potassium Dichromate) +/‐ 0.005 AU at 0.5AU (Nist 930e @ Visible Range) +/‐ 0.01 AU at 1.0AU(Nist 930e @ Visible range)

• Photometric Reproducibility: <0.002AU @ 0.5 and 1.0AU

• Stray Light: <1.0% (198 nm KCL), <0.05% (220 nm Nal), <0.03% (340 nm NaNO2)

• Photometric Noise: < 0.002 AU RMS @ 0 AU 300 nm

• Photometric Stability: < 0.002 AU/h @ 0 AU, 300 nm

• Interface: USB

• Environmental Requirements Temperature: 5‐40°C

• Environmental Requirements Humidity: up to 80%

• Non‐condensing

• Compact and modular unit provides ultrafast data processing and maximum reliability

• PDA (photodiode array) detector allows simultaneous data acquisition from 190 to 1100 nm wavelength

• Ideal for use in biotechnology, environmental, pharmaceutical QA/QC and academic labs

• Applications: QC/QA and R&D, kinetics

Specifications

Electrical Specifications: 120/240 V

Dimensions, Product: 13.4" W x 12.6" D x 4.5" H

Weight, Product: 14.6 lb

Dimensions, Shipping: 17158"W

Weight, Shipping: 32 lb

Package Type: Box

Unit of Measure: EA