Customer must determine if filter system leakage testing is required for their application. This unit is not full seam welded. If a leak test is required, full seam welded units leak-tested to meet ISO 14644-3 are optionally available.
Ultra-Low Particulate Air (ULPA) Filter: 99.999% efficient at 0.12µm particle diameter
Room-Side Replaceable (RSR) design allows quick filter replacement from the inside of the cleanroom. Because it eliminates the need for roof access or breaching the cleanroom, an RSR model greatly reduces maintenance time and costs.
Gel-sealed filter seats securely against a knife-edge on the fan housing
Easy-open clips allow fast filter replacement from the room side without breaching the cleanroom ceiling
Filter shipped uninstalled to prevent shifting in the mounting gel that could cause leaks.
AC 3-speed blower motor
FDA-approved silicone sealant
Internal baffling plate and diffuser panel provide quiet uniform air flow across the filter face
1/4-20 UNC threaded inserts on corners to allow suspension by guy-wires
MERV 7 pre-filter, 30% ASHRAE rated
Low-profile design is ideal for cleanrooms with limited ceiling space
Removable yellow power cord with 4-pin connector for easy installation to Terra cleanrooms (connector may be used with an optional wall plug adapter; or may be removed entirely to hard-wire the FFU)
Filters are leak tested in accordance with IEST standards and meet UL 900
See performance data and submittal sheets for additional information
Optional Accessories:
Challenge port kit
Power plug adapter
Stainless surface mount frame for installing FFU in ceiling without T-grid
12" ducted inlet connection adapter
Replacement filter
Filter replacement alarm
Eye bolt package
Ionizing bar
Pre-filter safety cover, for CE conformance
Control panel with night mode switch and power distribution modules
NOTE:
Customer must determine if filter leak testing is required for their application. This unit is not leak rated. If a leak test is required, units leak-tested to meet ISO 14644-3 are optionally available.
All filter sales are final; returns cannot be accepted due to the sensitive nature of HEPA filters. Leak tested filters may qualify for limited replacement if a defect is identified upon delivery.
For a full list of accessories, select a product then click Choose Accessories on that product page.
Night Service Saves Energy, Extends Filter Life!
Night service switch, shown mounted on Modular Hardwall Cleanroom panel, puts FFUs in low-energy mode during off-hours; saves 25% of fan operating costs and extends filter life.
Filter Replacement Alarm
The filter replacement alarm monitors filter backpressure and issues an alarm when filter needs replaced (requires selection of fan filter unit with challenge port, order separately).
Power Distribution Modules for Cleanroom FFUs and Lights
UL-listed power distribution modules include all breakers, transformers and other electrical components for FFU/light connections; allows fast addition or relocation of FFUs or lights without the need for rewiring.
Ionizing Bar; 22"L, with Installation Kit, for 24 x 24 Fan/Filter Unit
Ionizing bar, mounted to Fan/Filter Unit, provides a balanced stream of ions to the work area below to neutralize static charge and prevent electro-static discharge; 22" bar shown installed on 2' x 2' FFU.
Challenge Port; for Fan/Filter Unit
Challenge port for Fan/Filter Units simplifies filter leak testing and connection of Filter Backpressure Alarm (shown mounted on SS FFU, order separately).
Filter Covers Allow Ceiling Wash-Down
Ceiling panel covers allow wash-down and sterilization of cleanroom walls and ceilings without damage to filter medium; requires BioSafe glass or stainless steel wall panels, which stand up to biocides.
Cord Adapter, MIN4 PL to 16AWG Power Cord for Terra Fan/Filter Unit
Four-pin power cord adapter for fan/filter unit allows connection to standard 120VAC/60Hz power receptacle.
FFU Mounting Frame; Stainless Steel
Stainless steel Fan Filter Unit (FFU) mounting frame supports FFU on drywall or other rigid ceiling material, viewed from above ceiling panel (FFU not included).
Safely Mount FFU’s Above Floor for Horizontal Laminar Airflow
The portable FFU base stand can be easily positioned to supply Class 100 horizontal laminar flow into a room or enclosure. It can also be positioned next to sensitive equipment/materials that are mounted close to the floor.