This negative-pressure system is designed to ventilate fumes
to your in-house exhaust system, usually by means of an exhaust
duct.
If your facility includes a negative pressure exhaust system,
you can order a flange adapter panel for any of the Universal
Hoods, in a variety of materials and sizes (see below). This
panel provides a flange for connection of an exhaust duct
directly from the hood to your exhaust system.
Or, if no exhaust fan is present, order a blower module
out of either of the materials offered below. The blower
housing contains a 1200 CFM (2038.81 m³/h) impeller mounted
for negative-pressure applications. The blower housing
comes with a 8"-diameter x 1" (203 mm x 25 mm) high flange (made of the
same material as the housing you specify) for exhaust
ducting; other sizes may also be ordered. For twin stations, 72"W or 90"W (1829 mm or 2286 mm), two blower housings are generally required
for adequate exhaust venting.
Select Adapter Panel for connection to in-house ventilation
system.
|