Compliance:
  • SEFA 1-2005
  • ANSI Z9.5-2011
  • Modified ASHRAE 110-1995

Protector XVS Ventilation Stations are ideal for use as light-duty fume hoods, science classroom workstations, solvent cleaning work bays, weighing stations or for pathology and cytology applications. Tempered glass, 3/16" (5 mm) tops and sides allow full visibility
  • Low-profile, large capacity enclosure, 23"-deep (584 mm) interior for analytical work, solvent cleaning, pathology and cytology studies
  • By-pass airflow design maintains stable air velocity when air sash is opened or closed, eliminating high-velocity air streams that disturb delicate procedures
  • Patented Clean-Sweep™ air foil draws air under the air foil to minimize “dead air” zones and maximize containment, regardless of operator position
  • Patented Rear baffle with perforation zones pivots down and detaches for easy cleaning
  • Glacier white and gray, dry powder epoxy-coated aluminum frame, and steel rear plenum and baffle construction eliminates flimsy plastic-on-plastic bonding
  • Angle tempered glass sash pivots upward and locks for easy loading of large materials
  • Includes upper containment sash foil, upper dilution air supply, and side-entry air foils for optimal air velocity stability, two utility ports with iris openings and plastic plugs, top and bottom 2" x 10" (51 mm x 254 mm) exhaust collars for connection to FilterMate Exhauster and ductwork
  • Available Accessories: Work Surfaces, Base Stand with Leveling Feet, Guardian Jr. Airflow Monitor, Base Stand with Casters, Fluorescent Task Light, Corrosion Resistant Remote Blower, FilterMate Portable Exhausters

Protector XVS Ventilation Stations - Light Duty Fume Hoods

Labconco # Dimensions*
W" x D" x H"
(mm)
Total Exhaust CFM/Static Pressure**

Nominal

Width

Cat. # Price
80 fpm
(.4m/s)
S.P. 60 fpm
(.3m/s)
S.P.
4862010 24.0 x 29.0 x 22.8
(610 x 737 x 579)
95
(.48m/s)
.03"

70
(.36m/s)
.02"

2 ft. 3646-18 Price
4862030 24.0 x 29.0 x 32.0
(610 x 737 x 813)
120
(.6m/s)
.05"

90
(.46m/s)
.03"

2 ft. 3646-21 Price
4863010 36.0 x 29.0 x 22.8
(914 x 737 x 579)
145
(.73m/s)
.06"

110
(.56m/s)
.04"

3 ft. 3646-19 Price
4863030 36.0 x 29.0 x 32.0
(914 x 737 x 813)
185
(.97m/s)
.10"

140
(.7m/s)
.06"

3 ft. 3646-22 Price
4864010 48.0 x 29.0 x 22.8
(1219 x 737 x 579)
200
(1.0m/s)
.10"

150
(.76m/s)
.06"

4 ft. 3646-20 Price
4864030 48.0 x 29.0 x 32.0
(1219 x 737 x 813)
250
(1.2m/s)
.16"

185
(.94m/s)
.09"

4 ft. 3646-23 Price

* Internal Clearance: 21.6/33.6/45.6"W x 23"D x 20/29.3"H (549 mm/853 mm/1158 mm W x 584 mm D x 508/744 mm H ); Sash: 20° angle x 8" (203mm) opening, 22.8"H (579mm) models, 14° angle x 10" (254mm) opening, 32"H (813mm) models. Sash locking (open) position: 20"/29.25" (508 mm / 743mm) respectively.
** With sash closed, 8" (203mm) opening for 22.8"H (579mm) models, 10" (254mm) opening for 32"H (813mm) models

Video of Protector Fume Hoods by Labconco

Top of page

Accessories for Protector XVS Ventilation Stations

Ductless Exhaust Hoods Protector XVS Ventilation Stations from Labconco
Work Surface
For XVS Cat. # Labcono # Cat. # Price
2 ft. model 3646-18 3908400
3646-24 Price
2 ft. model 3646-21 3908400
3646-24 Price
3 ft. model 3646-19 3908401
3646-25 Price
3 ft. model 3646-22 3908401
3646-25 Price
4 ft. model 3646-20 3908402
3646-26 Price
4 ft. model 3646-23 3908402
3646-26 Price
† 2'/3'/4'W x 26.7"D (610 mm / 914 mm / 1219 mm x 678mm)

Base Stand with Leveling Feet
Durable 1.75" (44 mm) tubular, epoxy-coated steel construction and glacier white exterior. Telescoping legs are adjustable in 1" (25 mm) increments from 27.5" to 33.5" (698 mm to 851 mm). Includes leveling feet. Order work surface separately.
Quick Ship Item Items with prices in blue usually ship within 3 – 5 days of order.
  Labcono # Cat. # Price
24"W (610 mm)
3746700 3646-27 Price
36"W (914 mm)
3746701 3646-28 Price
48"W (1219 mm)
3746702
3646-29 Price

Base Stand with Casters (shown in photo above)
Epoxy-coated steel frame with glacier white exterior. Supports up to 1,000 lbs. (453.6 kg) load. Includes four 5" (127 mm) diameter toe-locking casters.
  Labcono # Cat. # Price
24"W (610 mm) 3746710 3646-80 Price
36"W (914 mm)
3746711 3646-81 Price
48"W (1219 mm)
3746712 3646-82 Price

Corrosion Resistant Remote Blower
Required when ducting to the outside. Direct-drive blower has .25 hp TEFC-type motor. Accepts 6" (152 mm) -dia. PVC duct.
  Labconco # Cat. # Price
115VAC, 60Hz 4863500 3646-32 Price
220VAC, 50Hz 4863501 3646-32-220 Price

FilterMate™ Portable Exhausters, from Labconco
  • Select HEPA filter, carbon filter, or combination of both to remove hazardous powders, particulates or vapors from the exhaust air stream
  • Converts the Protector™ XVS™ Ventilation Station into a ductless fume hood
 

Hoods — Guardian Jt. Airflow Monitor
Guardian Jr. Airflow Monitor
Top-mount unit, adds 6.4"H (163 mm) allows easy selection and calibration of a setpoint between 30 and 250 fpm (.15 m/s & 1.25 m/s). If hood’s face velocity falls below this level for more than six seconds, the unit issues an audible alarm sound and visible ALERT indication.
  Labcono # Cat. # Price
115VAC, 60Hz 3944700 3646-30 Price
220VAC, 50Hz 3944701 3646-30-220 Price

 


Hoods — Fluorescent Task Light
Fluorescent Task Light
  Labconco # Cat. # Price
115VAC, 60Hz 3909200 3646-31 Price
220VAC, 50Hz 3909201 3646-31-220 Price

 

Top of page

Compliance of Protector XVS Ventilation Stations

  • SEFA** 9-2010, DH I (when connected to a FilterMate Portable Exhauster)
  • SEFA** 1-2010 (Ducted models only)
  • SEFA** 8-2010, Cabinet Surface Finish Tests
  • Modified ASHRAE** 110-1995
  • ANSI** Z9.5-2011

View Other Ductless Exhaust Hoods and Enclosures Similar to the Protector XVS

Contact Terra Universal at 714-578-6100 for pricing and more information about applicable accessories. Click on the chat icon at the bottom of  the page and talk to a Product Specialist. 

Top of page

File Type File Name Publication Date
White Paper (PDF) Chemical Guide for Carbon Filtered Enclosures
Provides chemical specific data to assist in determining the suitability of ductless filtration for your application. Includes safety precautions when working with vaporous chemicals.
01/01/2015
Click to leave a message. 24-hour sales and tech phone support, M - F
Go to...
Go To
Laboratory-Equipment.com