 |
|
| |
Nitrogen Consumption |
Recovery Time |
| Fig. 1: |
110 ft3 (3.1 m³)
|
40 minutes |
|
Fig. 2:
|
55 ft3 (1.6 m³)
(100% improvement)
|
20 minutes
(100% improvement) |
|
Fig. 3:
|
15 ft3 (.42 m³)
(nearly 300% improvement)
|
10 minutes
(400% improvement) |
|
|
 |
 |
 |
Fig. 1:
Dual Purge™/
NitroWatch®Systems, One Automatic RB®
Valve
Moisture migrates through cabinet |
Fig. 2:
Dual Purge™/
NitroWatch®Systems,
Four Automatic RB®Valves
Moisture migration eliminated |
Fig. 3:
NitroPlex™ System
Efficient moisture removal,
reduced N2 use |
|
Performance Improvements Spell Big Savings:
NitroPlex™ Cuts N2 Costs by over 75%, Improves Efficiency by 400%
A desiccator with only one bleed valve may cause more harm than good.
Once an access door is opened, any moisture that enters a cabinet will tend to migrate from one chamber to the next until it exits the cabinet through a bleed valve. In a large, multi-chamber cabinet with only one bleed valve, the moisture may contaminate several chambers.
Even a desiccator equipped with Terra Dual Purge™ and NitroWatch® systems may require 40 minutes or more to purge the moisture and recover the desired humidity set point (see test results at right). In this test, the top chamber of a four-chamber desiccator was exposed to moisture until the humidity level reached 97%RH. Figure 1 (on graph) represents the desiccator efficiency when it was equipped with a single Automatic RB® Valve located in the bottom chamber. The system required 40 minutes to bring the humidity level down to the 10% set point.
If, during that period, another door is opened, this recovery time will be extended further. Very possibly, the cabinet might fail to reach its programmed RH set point over the course of an entire shift, or even longer.
Simply by installing Terra's Automatic RB® (Relief/Bleed) Valve in each desiccator chamber, you can cut both the system recovery time and the nitrogen consumption in half. This is the test case represented by Figure 2 on the graph. Only 20 minutes—and only half as much nitrogen--were required to bring the humidity level to the set point.
The final test case (Figure 3) included a NitroPlex™ control system. With this set up, the desiccator required only 10 minutes to bring the RH from 97% to the set point level of 10%—only a fourth the time required in the first test. It consumed only 15 cubic feet (.42 cubic meters) of nitrogen, only 13% of the amount used in the first test case.
This nitrogen saving alone, nearly 77% in this test, justifies investment in a NitroPlex control system.
|