Vertical laminar flow hood with top-mounted fan/filter unit.

Horizontal vs Vertical Laminar Flow Hoods

A laminar flow system is a vital concept in the control of particulate contamination. Laminar airflow is defined as air moving at the same speed and in the same direction, with no or minimal cross-over of air streams (or “lamina”). By contrast, turbulent flow creates swirls and eddies that deposit particles on surfaces randomly and unpredictably. Most contamination-sensitive environments such as dust-free hoods require laminar flow because it predictably sweeps particles in a uniform direction, from the cleanest area under the hood (the filter face) to the exit area, which is generally the sash opening or vents along the back or bottom of the hood. This design ensures that the cleanest (and most germ-free) area will always be the upstream area closest to the filter face. Work is generally done in that clean zone, as far as possible from obstructions that create turbulence.

Vertical laminar flow hood with top-mounted fan/filter unit.
Model shown includes rear exhaust baffles and a perforated base to optimize laminar flow, and an overhead duct to feed exhaust air to an in-house ventilation system.

But which direction should the laminar flow move: vertically (downward from a filter positioned above the work surface) or horizontally (forward from a filter positioned behind the work surface)? There are compelling reasons to choose either airflow system, depending on clearance requirements, process location, work surface design, and operator safety.

Vertical laminar flow hoods are often chosen because they resemble, on a small scale, the design of a laminar flow cleanroom, in which fan/filter units are typically positioned in the ceiling. By directing the laminar flow downward, vertical laminar flow reinforces the effect of gravity and sweeps particles out of the enclosure, generally through a front access area. Micro-contaminants may not have substantial mass, but most particles do eventually settle on a work surface or the floor of a room, and vertical flow helps get them there faster.

Vertical Laminar Flow Diagram

Room air (in red) is cleaned by the filter and forced in a downward direction toward the work surface.

Room air (in red) enters the system from above the HEPA filter; 99.99% particle-free air is forced downward toward the work surface.

A clean bench using a vertical laminar flow design is also preferable where floor space is at a premium. Because this configuration stacks the fan/filter module on top of the hood, a vertical laminar flow hood can generally be installed on a standard lab bench. By contrast, a horizontal-flow hood requires additional depth to accommodate the rear-mount filter/fan unit with some additional rear clearance to allow air to enter the system. These clearance requirements necessitate a deeper bench and more floor space than a vertical flow system.

Horizontal Laminar Flow Diagram


Room air (in red) enters the system from behind the HEPA filter; 99.99% particle-free air is forced in a back-to-front direction across the work surface.

Although both airflow designs provide effective sweeping action near the filter face, their respective flow patterns eventually encounter disruptions that often tip the scale in favor of one or the other configuration. In a vertical-flow hood, the obvious obstacle is the work surface. A perforated or rod-top work surface allows the laminar air stream to pass through the hood with minimal obstruction, but these tops can be a problem in operations involving liquids or small parts. If you’re working with these materials and want to avoid picking up parts off the floor, a horizontal flow design may be preferable.

Yet even a solid top may not rule out vertical flow if work is performed above the work surface. If sterile or particle-sensitive processes are performed in a clean, sterile zone midway between the work surface and the filter face, a vertical flow hood is generally acceptable. One such operation is sterile compounding, in which injectables or sterile packages are prepared above, not on, the work surface. As long as hands and other contamination sources move up and down, not sideways above a sample, sensitive materials will remain clean.

Finally, consider the effects on operators of air exiting the laminar flow hood. Although horizontal laminar flow, with air traveling from the rear of the hood and exiting the front opening, may not encounter large obstructions inside the hood, it does eventually encounter the person performing the work. Any annoying substances, such as soldering fumes or fine powders, may be blown into the operator’s face. While this collision may not compromise the laminar flow where work is performed, it may pose a health risk. In such cases, vertical flow is probably preferable.

Horizontal laminar flow hood, with rear-mounted fan/filter unit.

Horizontal laminar flow hood, with rear-mounted fan/filter unit.

Some final considerations involving both vertical and horizontal laminar hoods:

1) Neither vertical nor horizontal laminar flow hoods are appropriate in the presence of biohazards. Only an approved containment system, such as a Class II, Type A2 Biosafety Cabinet, should be used in applications requiring Biosafety Level (BSL) 2 or 3 containment.

2) Whichever laminar flow hood design you choose, be sure to run it for at least15 minutes before you begin critical operation, to allow particles to be purged from the enclosure.

3) Remember the sweeping direction, from clean to dirty, whenever you clean a hood. Wipe down surfaces in smooth, linear motions beginning at the filter face and moving downstream, with the air flow. Likewise, perform “dirty” operations, like unpackaging samples, as far from the filter as possible.

Vertical Laminar Flow Hoods


  • Hood not as deep: requires less floor space
  • Suitable for compounding sterile products
  • Safety: air not blowing directly at operator, and sash provides a barrier in front of operator’s face
  • Filter on top: easier to access
  • Less turbulent effect from air striking large objects or processing equipment
  • Less cross-contamination of items positioned on the work surface


  • Overhead clearance requirement; changing filters or servicing unit may require a step-ladder
  • Cannot place items or hands on top of other items: obstructs airflow
  • Increased turbulent effect of air striking work surface

Horizontal Laminar Flow Hoods


  • Reduced turbulent effect of air striking work surface
  • No sash: easier to work and position equipment, but air blows directly on operator
  • Hands and gloves are generally less contaminating since they’re downstream of the sample


  • Filter change or service usually requires repositioning hood for rear access
  • Large samples obstruct laminar air flow, may contaminate downstream samples
  • Blows fumes and/or powders in operator’s face
Terra Universal

Terra Universal

Terra Universal is the leading manufacturing expert in critical environment applications, with nearly 40 years of design and fabrication experience in cleanroom- and laboratory-based industries. Terra manufactures comprehensive lines of modular cleanrooms, pass-through chambers, glove boxes, laminar flow benches and furnishings. Custom work is no problem; our engineers work with yours to meet unique requirements.